Practical Tip | Aligning TEM Lamellae to the Zone Axis Without a Crystalline Substrate in the Region of Interest

TEM Lamellae to the Zone Axis
Listen to article
Practical Tip | TEM Lamella Zone-Axis Alignment for Semiconductor Analysis
4:45

TEM lamella alignment is essential for reliable transmission electron microscopy analysis, especially when the region of interest does not include a crystalline substrate that can guide zone-axis orientation. This workflow explains how an auxiliary silicon reference window can support accurate lamella orientation in semiconductor samples with tungsten contacts, gates, and dielectric layers.

Why Zone-Axis Alignment Matters in TEM Lamella Preparation 

In microelectronics, devices are typically built on a substrate with differently doped areas, overlayed with gates to control the current flow. Gates are then interconnected by multiple layers of metallization, separated by dielectric layers. A structural defect leads to device failure or affects performance. TEM lamella preparation makes it possible to investigate the region of interest at high resolution.

To obtain reliable TEM measurements, the lamella usually needs to be oriented to the zone axis, often using the monocrystalline substrate as a reference. This orientation is necessary to reach atomic-resolution imaging, accurate layer-thickness measurements, and clear layer contrast without imaging artifacts.

Image 1 shows a 30 keV STEM image of a TEM lamella from an Intel 32nm chip, including the Si substrate, gates, W contacts, and lines of the first metallization layer. In this example, the silicon substrate provides a clear reference for orienting the lamella in TEM.
  Image 1. SEM 30 keV STEM bright field image of a lamella from an Intel 32 nm chip at the gate level

Preparing Plan-View TEM Lamellae When the Defect Is Away From the Substrate

Sometimes, the expected defect is not located on the substrate. To localize it, a plain-view lamella parallel to the substrate must be prepared at a specific device level. Image 2 shows a TEM lamella prepared from a smartphone camera image sensor at the level of the gate contacts, which connects the first metallization layer with the gates and are surrounded by dielectric. At this point, there are two possible preparation tasks, depending on the expected location of the defect. In one case, the silicon substrate is removed completely from the back side, leaving only tungsten contacts and dielectric within the lamella. In that situation, no silicon remains available to orient the lamella. In the second case, when it is preferable to keep the gates and part of the substrate, the silicon substrate is thinned on the back side to approximately 100-200 nm.
  Image 2. SEM 1 keV SE image of the front side of the TEM lamella from a Xiaomi Mi5 Pro camera sensor at V0 level

Using an Auxiliary Silicon Window as a Zone-Axis Reference  

In this example, the lamella thickness is approximately 700 nm, including around 200 nm of remaining silicon and 500 nm of dielectric. As a rule of thumb, the amorphous layer in a lamella should not exceed 20% of the total lamella thickness. Traditionally, “amorphous” refers to the damaged layer induced by FIB polishing, but in this case, it refers to the amorphous dielectric layer. Although the high acceleration voltage of a TEM can still produce a good image of this lamella, it can be difficult to obtain a diffraction pattern from silicon when it is hidden by a large portion of dielectric. To provide a clear reference for orienting the lamella to the zone axis, a small area adjacent to the main thinned region can be used to open a second auxiliary window, which is polished until only the silicon substrate remains. Image 3 shows the final view of the lamella from Image 2, in which the leftmost 2 microns were polished further until they consisted solely of silicon, with a residual thickness of approximately 150 to 200 nm.
  Image 3. SEM 15 keV SE image of a TEM lamella from a Xiaomi Mi5 Pro camera sensor at V0 level 

Verifying Lamella Orientation With Diffraction and STEM Imaging 

Image 4 shows an overview bright field STEM image of the lamella in TENSOR with its diffraction image. The diffraction image indicates that the lamella is correctly oriented in the zone axis using a silicon substrate. Image 5 shows a STEM bright field image of a lamella detail after alignment in the zone axis. Together, these views show how the reference window supports zone-axis alignment when the region of interest does not provide a clear crystalline reference. 

STEM-imaging

Written by Maksym Klymov
Head of Applications Department, Tescan  

Recent Posts

More TEM Preparation Resources

TEM AutoPrep™ Pro
Automated TEM lamella preparation with FIB-SEM.

Explore the Solution

Automated TEM Sample Preparation
Automated workflows for precise TEM lamella preparation.

Automated TEM sample preparation workflows

Get in touch

map

Your nearest office is:

Tescan Brno
Libušina tř. 21
623 00 Brno - Kohoutovice

Czech Republic

+420 530 353 411
info@Tescan.com

130405923 us US 37.09024 -95.712891 25.3575 29.349345 20.67957527 42.082797 39.91384763 -33.693421 13.93320106 3.039986586 31.997988 38.050985 47.579533 48.1485965 58.375799 54.663142 19.195447 56.975106 47.916997 50.493053 45.868592 10.79556993 44.35660598 43.2371604 55.536415 14.557577179752773 32.100937 -6.116829 -6.212299277967318 33.600194 -12.08688 23.7104 -33.471062 31.998740087 -23.69149395 43.462349 51.529848 49.1893523 49.197486 25.072375 31.075811 1.299027 40.676979 52.30150662 51.013813 35.684121 37.566531 52.246622 40.581349 39.911632 -26.1811371 41.818215 33.429928

No distributors found.